Remote mode is utilized for EPS control of the turbo pump, but we only utilize 4 DI and 1 DO functions (pin 3, 4, 5, 8, and 14) and only 4 EPS PVs to control (not 5).
DI remote priority on pin 14 (control via the remote interface - 26-pin D-sub):
open --> off
V+ --> set and priority over other digital inputs.
EPS group is looking into the EPS side (as there isn't currently an EPS PV to change this)
In short term, add to the RS-485 ioc the ability to read remote priority. Is this a global DI and DO?
Also add to the RS-485 ioc "Control via the interface" using parameters 060 and 061.
Remote mode is utilized for EPS control of the turbo pump, but we only utilize 4 DI and 1 DO functions (pin 3, 4, 5, 8, and 14) and only 4 EPS PVs to control (not 5).
DI remote priority on pin 14 (control via the remote interface - 26-pin D-sub):
In short term, add to the RS-485 ioc the ability to read remote priority. Is this a global DI and DO?
Also add to the RS-485 ioc "Control via the interface" using parameters 060 and 061.
http://www.idealvac.com/files/ManualsII/Pfeiffer_TC400_Operating_Instructions.pdf