Timebase seems wrong #30
Comments
Hi @CM630 , Does it happen on 10ms timebase only? |
It happens on all timebases.
|
I am not seeing this on my hardware. It is possible that your hardware revision changed (maybe has a faster processor). |
I have tried with all 4 channels enables - still the timebase is wrong. |
On DLO-138_switches_1.0.bin from 26 Feb 2017:
When the indication is 10 ms/div it is actually 5 ms/div.
So a pulse with duration 10 ms shall be displayed in a single section.
When I supply 50 Hz to the input of the oscilloscope at 10 ms I shall see 6 periods (12 section* 10 ms = 120 ms= 6 periods).
Instead, I see 3 periods only.
For example: https://github.com/ardyesp/DLO-138/raw/master/pics/pic4.png (probably made on another version of the software) shows 2 ms per division chosen, and about 1 period of 60 Hz visualized, but with the aforementioned version I see about half a period at 50 Hz.
Also, when a waveform is sent over UART, the value of „Per Sample (us):“ and „Timebase: 10 mS/div“ seem wrong, the Time column- too.