You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
A forward slash or backslash in the filename crashes the CDSAXS driver (eek!). This is deep and subtle. It happens in the sub-VI running on the Pilatus server, the count never ends and the main labview becomes unresponsive.
Thought this was sanitized, but evidently not.
The filename that broke it was : "H2O blank 6.00 milligram / milliliter 1000.00 mgml"
--PAB
The text was updated successfully, but these errors were encountered:
To fix: connect to the labview Pilatus server @ 192.168.0.200 via RDP and stop the counter vi there. That should restore your access to the main panel.
A forward slash or backslash in the filename crashes the CDSAXS driver (eek!). This is deep and subtle. It happens in the sub-VI running on the Pilatus server, the count never ends and the main labview becomes unresponsive.
Thought this was sanitized, but evidently not.
The filename that broke it was : "H2O blank 6.00 milligram / milliliter 1000.00 mgml"
--PAB
The text was updated successfully, but these errors were encountered: